Measuring Samples

Thermo Scientific
Parameter
Description
Delay Time
Enters the time from Test Initiation to first measurement; allows for
sample equilibration
Adv.
Interval Time
Enters the time between repeated readings
Measure Blank
Selects the frequency of zeroing the instrument
(as function key)
Once or Every Reading
2. When the parameters are set, press Save Test to save the test or Run Test to measure a
blank or sample.
Note If Cell Correction is ON, you must run the Setup Correction application before
you can access Run Test or Measure Samples.
Note If Auto Save Data is ON, you must enter a Data File Name before you can
access Run Test or Measure Samples.
Note If a 6-Position Cell Changer is installed, place the blank in position B and the
sample in cell position #1. The instrument always uses cell position #1 to scan the sample.
To measure samples
1. In the Kinetics screen, press Run Test.
2. If a 6-Position Cell Holder is installed, place the blank in position B and the sample in
position #1.
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
21
Kinetics
Measuring Samples
101