Specifying Wavelengths for Discrete nms Mode
Thermo Scientific
7. Press Run Corr. to start Cell Correction.
If you selected Discrete nms mode, first specify the wavelengths using the following
procedures, and then Cell Correction.
Cell Correction measures the other cells against the blank and records, stores and dates
the measurements. From these measurements Cell Correction establishes the required
correction factors, which then are automatically applied during all subsequent tests (if
Cell Correction is activated).
To specify wavelengths for Discrete nms mode
1. Highlight Sample Positioner and press Enter to set this parameter to either Auto 3
(when using three large cell holders) or Auto 6 (when using six small cell holders).
2. Highlight Number of Matched Cuvettes and press Enter.
3. Specify the number of cells you are matching and press Enter.
4. Press Set nms to select the wavelengths for which Cell Correction will be run.
A list of wavelengths appears.
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
5
Cell Correction
Cell Correction
23