Maintenance

Changing the Fuse

Cleaning the Windows of the Sample Compartment

Changing the Fuse
122
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
IMPORTANT Do not dry cells in an oven.
Micro flowcells can be kept clean by:
• Flushing well with a solvent after use.
• Aspirating dilute acid, base, non-filming detergent or Clorox® through the cell in short
bursts.
• Storing with distilled water in the cell.
Do not use acetone or abrasive materials to clean the windows of the sample compartment.
Instead, use a non-abrasive laboratory cleaning solution (such as a commercial cell cleaning
solution), distilled water or alcohol.
Use the liquid and a soft, lint-free cloth to clean the windows. Do not apply too much
pressure or the surface of the windows may be damaged. Be sure to remove all fingerprints.
The fuse is located in the power entry module located at the center of the back panel of the
instrument.
• 120 VAC, 2.5 A, Slo Blo
• 240 VAC, 1.25 A, Slo Blo (2 required)
IMPORTANT The instrument fuse must be replaced with the same type and rate.
IMPORTANT If the fuse fails repeatedly, it may indicate a serious problem with the
instrument. Contact technical support as soon as possible.
To change the fuse
1. Turn off and unplug the instrument from the wall outlet or power strip.
2. Position the instrument so you can access the power entry module on the back of the
instrument.
3. Remove the power cord.
4. Insert a flat-blade screwdriver into the notch on the fuse cover and pry off the cover.