22
Performance Verification

Troubleshooting Checklist

Troubleshooting Checklist
108
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
If a Performance Verification test fails, follow the instructions below to diagnose common
problems.
If a test continues to fail after you have tried all these recommendations, follow the
troubleshooting list for the test being run (included with the description of each test).
Make sure:
• You follow the instructions for the test properly.
• Filters and standards are clean.
• The sample compartment door is closed during the test.
• The sample compartment is clear of obstructions.
• The cell holder assembly is installed properly. If the 6-Position Cell Holder is installed,
run the test once with the sample compartment door open to verify that the 6-Position
Cell Holder is moving smoothly.
• No problems are indicated by the power-on diagnostics after you turn the instrument
power OFF and then ON.
• The lamp is ON.
• The lamp compartment is clear of obstructions.
WARNING Do not open the lamp compartment unless the instrument power is OFF.
WARNING Do not turn the instrument power ON unless the lamp compartment is
closed.