Performance Verification

Accessing the Performance Verification Tests

Thermo Scientific
Performance Verification lets you check the performance of your instrument with these tests:
Wavelength Accuracy - Internal
Wavelength Accuracy - Standard
Wavelength Repeatability (UV-Vis only)
Resolution (UV-Vis only)
Photometric Accuracy
Noise
Stray Light
Run the appropriate tests regularly and maintain a log of results to help document the
reliability of the instrument and indicate potential performance issues.
Note If a printer is installed and turned on, the instrument automatically prints test
results. You can also press Print to print another copy of the results.
To access the Performance Verification tests
1. Press Tests.
2. Highlight Performance Verification and press Enter.