Cell Correction

Cell Correction
Thermo Scientific
Every test setup screen provides access to Cell Correction.
Note Cell Correction is not active from the Main (Basic ATC) screen.
Note Cell Correction is active only when the 6-Position Cell Holder is set to either
Auto 6 or Auto 3. The feature is not active when the cell holder is set to 1-Cell or Manual
6, nor when the Single Cell Holder is installed.
Before running Cell Correction:
• Clean the inside and outside of all the cells to be matched.
• Fill the cells with distilled water (or other blank solution), and place them in the sample
compartment (see
"Selecting and Positioning Cuvettes"
cuvette in Cell "B" of the cell changer.
To run Cell Correction
1. Load the test type or stored test.
2. If Cell Correction is not visible, highlight More parameters and press Enter.
on
page
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
5
5). Place the blank
21