Performance Verification
Photometric Accuracy
Selecting the Mode
Adding Standards
Deleting Standards
114
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
To switch between absorbance and %Transmittance, press Change to %T (or Change to
ABS) until the appropriate mode appears.
You will need to set three values whenever you add a standard: the wavelength, the absorbance
(or %Transmittance) and the tolerance value.
To add a standard
1. Press Add Std and enter the wavelength value in the Entry field.
2. Press Enter or Add nm to add the wavelength to the list.
3. Enter the absorbance or %T value for the entered wavelength in the Entry field.
4. Press Enter.
5. Enter the tolerance for the entered wavelength in the Entry field.
6. Press Enter.
The test screen displays the values you just entered for that standard.
7. Press Start Test to begin the measurement or Press Esc to save the test.
To delete a standard
1. Highlight the appropriate standard.