Editing a Standard Curve

Thermo Scientific
To measure samples manually (using Manual 6 or Single Cell Holder)
1. Press Run Test.
2. Install the blank and samples.
3. Press Measure Blank.
If a 6-Position Cell Holder is installed, it moves to the B position to measure the blank
and returns to its previous position.
4. Press Measure Samples.
If a 6-Position Cell Holder is installed, press the cell position buttons to reposition the cell
holder and measure the rest of the samples manually.
When the instrument has measured all the samples, the Standards screen shows the
absorbance and concentration of each sample.
You may edit the concentration of any standard on a standard curve. In addition, you may
change the number of standards, select a different curve fit or delete points from the curve.
20
Concentration Measurements—Standard Curve Application
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
Editing a Standard Curve