Wavelength Repeatability (UV-Vis only)

Thermo Scientific
2. Press Delete nm.
To run the test
1. Verify that the wavelengths and tolerances are set correctly.
2. Press Start Test.
The results indicate pass or fail for each wavelength.
If the test fails, follow these guidelines:
• Repeat the test twice to verify that the test is failing consistently.
• Make sure the target and tolerance values you entered for the calibrated wavelength are
the same as the values on the calibration certificate for the standard.
This test measures the ability of the spectrophotometer to return to an identical wavelength in
a repeatable manner. The test uses the internal xenon lamp.
A xenon lamp has a strong, fundamental line 529 nm. This line is an essential property of
xenon and serves as a fundamental standard.
When running the internal standard test, remember that:
• The wavelength and tolerance values are preset and cannot be changed.
• The cell holder should be empty.
To run the Wavelength Repeatability test
1. Highlight Wavelength Repeatability and press Enter.
2. Press Start Test.
The results indicate pass or fail for each wavelength.
Wavelength Repeatability (UV-Vis only)
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
22
Performance Verification
111