Concentration Measurements—Standard Curve Application
Recalling a Standard Curve
Recalling a Standard Curve
Setting the Parameters for a Standard Curve
Measuring the Standards for a Standard Curve
92
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
To recall a standard curve
1. Press Stored Tests.
2. Highlight the test to recall and press Enter.
To set the parameters for a standard curve
1. Place the standards in their cell positions.
2. Set parameters for measuring the standards.
See
"Parameters"
on
a. Enter the Test Name, Wavelength, Reference Wavelength Correction and
Reference Wavelength.
b. Select the Curve Fit, Units and Sample Positioner.
c. Set Number of Standards and Number of Samples.
d. Enter the low and high limits.
e. Select the settings for Statistics and AutoPrint.
f.
Run Cell Correction.
To measure standards automatically (using Auto 6 or Auto 3)
1. Install the blank and standards.
2. When the parameters are correct, press Run Standards.
page 125
for a complete list.