Photometric Accuracy
Thermo Scientific
2. Press Start Test.
The results indicate pass or fail for each wavelength.
If the test fails, follow these guidelines:
• Repeat the test twice to verify that the test is failing consistently.
• Contact technical support for more troubleshooting advice.
This test measures the absorbance (or %T) of a set of standards and compares the results with
specified tolerances. The wavelength absorbencies and tolerances are preset, but you should
change them to the values on the calibration certificate included with your standards.
Note You can display the tolerances for this test in either absorbance or %Transmittance.
When running the Photometric Accuracy test, remember:
• You need Photometric Accuracy standards calibrated to known absorbance values at
specified wavelengths.
• Measure the standards in the order they are listed on the test screen.
• You can use one to five standards.
To display the Photometric Accuracy screen
1. Highlight Photometric Accuracy.
2. Press Enter.
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
22
Performance Verification
Photometric Accuracy
113