Entering a Factor

Measuring Samples

Thermo Scientific
4. Insert the blank and standard.
5. Press Enter to measure the blank and samples and display the absorbance and calculated
factor.
To measure a standard manually (using Manual 6 or Single Cell Holder)
1. Press Run Standard.
2. Enter the concentration of the standard and press Enter.
3. Insert the blank and standard.
4. Press Measure Blank.
If a 6-Position Cell Holder is installed, it moves to the B position to measure the blank
and returns to its previous position.
5. Press Measure Standard.
The instrument measures the absorbance of the standard and displays the absorbance and
calculated factor.
To enter a factor
Highlight Factor. To change the factor, enter the correct factor.
To change the units, highlight Units and select the correct units.
To measure a sample automatically (using Auto 6 or Auto 3)
1. Press Run Test.
2. When prompted, place the blank and samples in their cell positions and press Enter.
The instrument measures the blank and samples and displays the sample measurements.
3. Press Measure Sample to measure additional samples.
To measure samples manually (using Manual 6 or Single Cell Holder)
1. Press Run Test.
2. Insert the blank and sample.
A 6-Position Cell Holder can hold five samples.