Remove The Rear Housing - Welch Allyn Connex Vital Signs Monitor 6000 Series Service Manual

Hide thumbs Also See for Connex Vital Signs Monitor 6000 Series:
Table of Contents

Advertisement

Service manual

Remove the rear housing

No.
1
2a
3a
6
Before you begin, power down the device, disconnect the AC power cord, and remove
the battery as described in this manual.
1. Lay the monitor face down on the antistatic mat.
2. Loosen the screw (7) to open the communications door (6).
3. Open the communications door (6) to access the rear housing screw.
4. Remove the four rear housing screws (14).
5. Stand the monitor on its feet with the display facing away from you.
6. Pull the top of the rear housing away from the monitor to access the rear housing
cable connections.
7. While supporting the rear housing, disconnect the following from the
communications module:
Item
Rear housing
USB label 4 USB host
Nurse call label
Communications door
No.
Item
7
Communications door captive screw
8
IEC connector
14
Screw, M4 X 10 pan head with NYLOC
Section B 127