104 Section A
Install the USB cables based on the type of cable separator you have—6-hole or 11-
hole—and the instructions below.
Note
USB cable locations—6-hole cable separator
Place the USB cables in the cable separator.
Note
Note
Note
Position each cable to extend approximately 1.5 inches past the separator.
Place the separator in the unit as shown below.
Plug the USB cables into the main board as follows:
The smart-battery harness to J34 on the main board.
The printer harness to J17 on the main board.
The speaker cable on the main harness to J12 on the main board.
First inspect J12 for bent pins before installing the
Note
speaker cable.
The 6-hole separator has been discontinued, but it might be
found in older devices.
If the USB cable was not removed from the temperature
module, leave the J1 position in the figure below vacant until
the cable separator is placed in the case and the other USB
cables have been connected to the main board. When the
cable separator is in the case, connect the USB cable from
the temperature module to J1 on the main board and snap
the cable into the J1 position in the cable separator.
The cable with right-angle connectors at both ends is in the
J14 position in the figure below.
If the option is not included, leave the separator position
empty.
®