112 Section A
4. Disconnect the light-bar harness from J46 on the main board.
5. Remove the light-bar board (34) and light bar (33) from the front housing.
6. Remove the light-bar board from the light bar.
7. Disconnect the light-bar harness from the light-bar board.
Reassembly notes
Insert the light-bar board into the light bar as follows:
1. Place the end without the connector into the closed end of the light bar, with the
2. Ensure that you slide the non-connector end of the light-bar board to the end of
Using the end furthest from the ferrite bead, plug the light-bar harness into the light-
bar board.
Route the light-bar harness as follows:
Secure the light-bar harness with tape, making sure that the ferrite bead is between
the main board and the tape.
Remove the temperature module or housing
Note
Note
Before you begin, remove the following as described in this manual:
Battery
Rear housing
CO2 module or blank faceplate
EarlySense module
SpO2 module or blank faceplate
LEDs facing into the light bar.
the slot in the light bar.
Through the slot in the front housing, underneath the horizontal strut.
In the channel and between the top of the housing and the LCD frame standoffs.
These instructions also apply to removing the temperature housing for
models without the temperature option.
Do not disconnect the USB cable from the temperature module unless you
plan to replace the module and/or the USB cable. Disconnect the USB
cable at the main board when removing the module.
®