1. Disconnect the LCD harness from the LCD. For easier access to the connector, you
can slide the ferrite bead closest to the connector up the harness.
2. Remove the four shoulder screws that secure the LCD frame.
3. Remove the LCD frame.
Note
4. Remove the LCD.
Reassembly notes
Verify that the LCD flex cable feeds through the bezel's clearance feature. Insert the
LCD into the bezel in the front housing with the LCD flex cable on the lower right as
shown. (The image on the left applies to 1st generation displays while the one on
the right applies to 2nd generation displays.)
1st generation displays
Place the LCD frame over the LCD. Verify that the LCD frame does not cover the
LCD flex cable.
It is not necessary to remove the temperature module or light bar to
remove the main board or LCD.
Be sure to support the LCD if you are removing it with the
frame. The foam strips on the LCD might cause the LCD to
stick to the frame.
CAUTION The solder joint of the LCD connector (J48) is fragile. When
inserting or removing the flex cable, do not stress the solder joint.
2nd generation displays
Section A 121