176 Section B
7. Insert the temperature module into the front housing aligning the probe well with the
access hole in the front housing.
8. Secure the top of the temperature module to the front housing with two screws
included with the kit.
9. Place the HSG clamp on the bottom of the temperature module and secure with two
screws.
10. Connect the USB cable to the module as described in the instructions for replacing
the temperature module.
Remove the main board
Note
Note
No.
10
42
46b
47
48b
If you are replacing this board, you must have the service tool, Gold
licensed edition, to re-enter the serial number of the monitor and the main
board.
If you are replacing the display, you must replace the bezel, LCD with
touchscreen, and foam pads with the replacement parts in Service kit
106825. The bezel and foam pads are designed to match the display.
Item
Screw, plastite #4-20 X 0.500 Torx-pan head
Front housing
Main board
Screw, M3 x 5 pan head
LCD bezel
®
Welch Allyn Connex
Vital Signs Monitor 6000 Series™
No.
Item
51
Foam pads, top and bottom
62b
LCD frame
63
Grommet, ear - G411-1
64
Shoulder screw, ear G-411-1 metric
95
Ground wire