Message
Cannot analyze ECG.
Cannot measure ECG.
Searching for respiration.
Possible cause
ECG module has not transmitted ECG
data for the past 30 seconds
ECG module disconnected
Unexpected alarm compatibility bits
Module unable to analyze the ECG
signal for V-Tach, V-Fib, and/or asystole
ECG module has been unable to detect
and ECG waveform for the past 30
seconds
ECG unable to measure patient
respirations
Troubleshooting 57
Suggested action
the ECG module if
necessary.
Check the USB cable and
connections.
Perform functional test. If
the test fails, contact
Customer Care and replace
the ECG module if
necessary.
Check the ECG module
connection to Connex
device.
Restart the Connex device
and the ECG module; if the
error persists, contact
Customer Care.
Perform functional test. If
the test fails, contact
Customer Care and replace
the ECG module if
necessary.
Check electrodes and
leads, and replace as
needed.
Check the patient cable
connection to the ECG
module.
Perform functional test. If
the test fails, contact
Customer Care and replace
the ECG module if
necessary.
Check electrodes and
leads, and replace as
needed.
Check the patient cable
connection to the ECG
module.
Perform functional test. If
the test fails, contact
Customer Care and replace
the ECG module if
necessary.
Check electrodes and
leads, and replace as
needed.