Housing top extension (extended chassis only)
No.
Item
74
M4 standoff
75
Housing top extension
Before you begin, remove the following as described in this manual:
Battery
Rear housing
CO2 module or blank faceplate
EarlySense module
SpO2 module or blank faceplate
NIBP module
Bottom housing
Note
1. If your model has an extended chassis, remove the housing top extension.
a. Use a 5/16-inch socket and driver to remove the two standoffs (74) that secure
b. Remove the housing top extension.
2. Lay the main harness over the top of the unit to clear the USB cables.
3. Disconnect the power connector on the main harness from J30 on the main board.
Do not remove the side panel that contains the power button.
the housing top extension (75).
No.
Item
79
Top housing gasket
Section A 101