166 Section B
Position each cable to extend approximately 2 inches past the separator.
When positioning the separator in the device, place the feet in the marked areas of
the MCE board as shown.
Plug the USB cables into the main board as follows:
No.
Item
J1
Temperature
J2
Printer
J3
EarlySense
J21
CO2
J24
CO2
For models with a temperature module, if you disconnected the USB cable from the
temperature module during disassembly, you must replace the retaining clip. (See
the topic "When replacing the temperature module" for specific instructions.)
For models with a printer, plug the other end of the USB cable into the printer while
pulling the plastic latch away from the connector. Once the connector is seated,
push the plastic latch over the connector to secure it.
®
Welch Allyn Connex
Vital Signs Monitor 6000 Series™
J1 J2
J3
J21
J24
J23
J26
No.
Item
J23
SpO2
J26
NIBP
J8
COMMS J4
J14
COMMS J8
J8 J14