Welch Allyn Connex Vital Signs Monitor 6000 Series Service Manual page 57

Hide thumbs Also See for Connex Vital Signs Monitor 6000 Series:
Table of Contents

Advertisement

Service manual
NIBP messages
Message
Alarm
NIBP air leak; check cuff and tubing
connections.
NIBP not functional. Call for
service.
Unable to determine NIBP; check
connections; limit patient
movement.
Unable to determine NIBP; check
connections and tubing for kinks.
Incorrect NIBP cuff size; check
patient type.
Inflation too quick; check NIBP cuff
and tubing connections.
Unable to determine NIBP; check
inflation settings.
Information
User cancelled NIBP reading.
Possible cause
The NIBP module has an air leak
Internal errors or messaging errors occurred
The ambient temperature is out of range
Pressure exceeded the maximum limit for this
patient mode
NIBP tubing on the outside of the device is kinked Check the connections and tubing
The NIBP module needs to be calibrated
Tubing inside the NIBP module is kinked
A neonate cuff is in use with the monitor in adult
or pediatric mode
NIBP inflation was too quick
Target pressure was too low
Too many attempts
Blood pressure reading cancelled by user
Troubleshooting 51
Suggested action
Check the cuff, tubing, and
connections for leaks. If no leaks
are found, retake the
measurement. If the message
reappears, replace the NIBP
module.
Check the error logs for NIBP
errors. See the service tool help
files for details on specific errors
and suggested actions.
Check with Welch Allyn for
software updates.
Use the monitor in the specified
temperature range.
Check connections; limit patient
movement.
Clear the alarm and retry NIBP.
for kinks.
Clear the alarm and retry NIBP.
Perform NIBP calibration.
Replace the NIBP module.
Check the patient type and cuff
size.
Clear the alarm and retry NIBP.
Check the connections and tubing
for kinks.
Clear the alarm and retry NIBP.
Check inflation settings and
change as necessary.
Clear the alarm and retry NIBP.
Change the inflation setting.
Change the inflation setting.
OK
Touch
to dismiss.