Verify that the Ethernet cable passes freely through the space between the
temperature housing and the printer board, exiting the case in the space between
the top of the temperature housing and beneath the top housing.
If the Ethernet cable was removed, re-install it when re-installing the bottom
housing.
When installing the top housing on the chassis, line up the slots on the top housing
with the ends of the horizontal struts on the chassis to ensure proper seating.
When installing the housing top extension (75), verify that the top housing gasket is
in place. Use a 5/16-inch socket and Torque driver calibrated for 12.0 inch-pound
±1.0 inch-pound to secure the standoffs (74).
When replacing the top housing
For models without a printer, install the printer door blank as follows:
1. Remove the paper backing from the gasket on the door blank.
2. Install the door blank into the top housing, pressing to secure the door blank to the
top housing.
Remove the printer
Before you begin, remove the following as described in this manual:
Battery
Rear housing
CO2 module or blank faceplate
EarlySense module
SpO2 module or blank faceplate
NIBP module
Bottom housing
Top housing
Section B 167