Welch Allyn Connex Vital Signs Monitor 6000 Series Service Manual page 53

Hide thumbs Also See for Connex Vital Signs Monitor 6000 Series:
Table of Contents

Advertisement

Service manual
When the device detects certain events, a message appears in the Device Status area at
the top of the screen. Message types include the following:
Information messages appear on a blue background.
Very low-priority alarms appear on a cyan background.
Low- or medium-priority alarms appear on an amber background.
High-priority alarms appear on a red background.
Technical alarm messages are low priority or very low priority unless noted in the
Message column.
You can dismiss a message by touching the message on the screen, or, for some
messages, you can wait for the message to time out.
To use these tables, locate the message that appears on the device in the left column of
the table. The remainder of the row explains possible causes and suggests actions that
can resolve the issue.
If you cannot resolve the issue, use the service tool to read the error log files or use the
service tool to perform a functional test
CO2 messages
Message
CO2 not functional. Call for service. A nonrecoverable communications error has
Filter line disconnected.
Occlusion in gas line. (High-priority
alarm)
CO2 temperature out of range. CO2
is not calibrated.
1
Requires the service tool, Gold licensed edition.
1
on the module reporting the message.
Possible cause
occurred
The sampling line is not connected to the monitor Connect a sampling line to the
The sampling line is defective
The input connector is dirty
The sampling line detection switch is faulty
The purge cycle failed to clear a blockage in the
sampling line
The purge cycle failed to clear a blockage inside
the module
CO2 module temperature is out of range
Troubleshooting 47
Suggested action
Check the error logs for CO2
errors. See the service tool help
files for details on specific errors
and suggested actions.
Check with Welch Allyn for
software updates.
monitor.
Check to ensure that the
sampling line plug is tightly
connected to the monitor.
Replace the sampling line.
Clean the input connector. See
the appendices for instructions.
Replace the module.
Replace the sampling line.
Replace the module.
Wait for the module temperature
to return to normal and the