184 Functional verification and calibration
To use the Standard unlicensed edition to test the NIBP module, follow the
instructions in this service manual.
For instructions on using the Gold licensed edition, see the service tool help files.
Standard unlicensed edition service tool tests performed
The service tool performs the following functions on the host device and installed
options.
Test
Reads
firmware
Firmware
upgrade
Leak test
Accuracy test
(NIBP)
Overpressure
test
Gold licensed edition service tool tests performed
The service tool tests the host device and installed options as listed in the following
table.
Test
POST
Firmware
version
Firmware
upgrade
Leak
AD noise
Calibration
Description
NIBP
Displays the
firmware version
Loads the latest
firmware into the
module
Verifies leaks using
100 cc volume
Checks the
accuracy of
transducers across
the pressure range
Verifies the
maximum cuff
pressure for adults
and neonates
Description
NIBP Temp SpO2 SpHb RRa
Performs the power-on self
test (POST)¹
Checks the firmware
version
Loads the latest firmware
into the module
Verifies leaks using 100 cc
volume
Checks noise on the
pressure channel
Calibrates pressure
transducers
®