174 Section B
Route the power button flex cable up and over the temperature module and plug the
cable into the J6 ZIF locking connector on the main board.
When replacing or adding a temperature module
Follow these instructions when replacing temperature modules and when installing them
in the device for the first time.
1. Prepare the temperature module housing: Attach the cable tie mount (113P463) with
a .75-inch square foam strip (78P567) to the SureTemp housing near the top corner
of the housing as shown.
2. Connect the USB cable to the temperature module.
3. Position the retaining clip on the USB cable against the shoulder of the mini-B
connector with the adhesive pad facing away from the temperature module and
toward the bottom.
4. Remove the backing on the retaining clip to expose the adhesive.
5. Rotate the clip on the USB cable to fasten the clip to the temperature module
housing.
®