288 Appendices
Calibration key temperature test for SureTemp Plus
Temperature test
97.3 °F (36.3°C)
SureTemp Plus test
9600 Plus Calibration Tester ID number
Temperature
tested
96.8 °F (36.0 °C)
101.3 °F (38.5 °C)
105.8 °F (41.0 °C)
Braun ThermoScan PRO 4000 or PRO 6000 test
9600 Plus Calibration Tester ID number
Temperature tested
96.8 °F (36.0 °C)
101.3 °F (38.5 °C)
105.8 °F (41.0 °C)
Specification ± 0.2°F (± 0.1°C)
97.1 to 97.5°F
36.2 to 36.4°C
Specification
Range
± 0.3 °F (± 0.2 °C)
96.5 °F to 97.1
°F
± 0.2 °F (± 0.1 °C)
101.1 °F to
101.5 °F
± 0.3 °F (± 0.2 °C)
105.5 °F to
106.1 °F
Specification ± 0.4 °F (± 0.2 °C)
96.4 °F to 97.2 °F
35.8 °C to 36.2 °C
100.9 °F to 101.7 °F 38.3 °C to 38.7 °C
105.4 °F to 106.2 °F 40.8 °C to 41.2 °C
®
Welch Allyn Connex
Vital Signs Monitor 6000 Series™
Actual reading
°F or °C
Calibration Date
Actual reading
35.8 °C to 36.2
°C
38.4 °C to 38.6
°C
40.8 °C to 41.2
°C
Calibration Date
Actual reading
°F or °C
°F or °C
°F or °C
Pass
Fail
Next Calibration Due Date
Pass
Fail
°F or
°C
°F or
°C
°F or
°C
Next Calibration Due Date
Pass
Fail