Service manual
Reassembly notes
Install the NIBP module (the module closest to the display) before installing the
SpO2 module.
Insert the SpO2 module into the housing with the serial number label facing up.
When replacing the SpO2 module or adding an SpO2 module
On the back of the SpO2 module, clean the surface to which the retaining clip will be
applied (see figure below) and apply the clip.
Insert the SpO2 module into the housing with the serial number label facing up.
Connect the USB cable to the SpO2 module.
Secure the USB cable to the retaining clip on the SpO2 module.
Remove the NIBP module
Before you begin, remove the following as described in this manual:
Battery
Rear housing
CO2 module or blank faceplate
EarlySense module (optional)
SpO2 module or blank faceplate
1. Disconnect the USB cable from the NIBP module.
CAUTION Ensure that your module has the correct option (Masimo or
Nellcor).