Welch Allyn Connex Vital Signs Monitor 6000 Series Service Manual page 119

Hide thumbs Also See for Connex Vital Signs Monitor 6000 Series:
Table of Contents

Advertisement

Service manual
NIBP module
Bottom housing
Top housing
Note
1. Remove the probe well from the temperature module.
2. Disconnect the power button flex cable from the J6 ZIF locking connector on the
main board and remove the securing tape.
3. Remove the right side panel (37).
4. Remove the two Torx screws (10) from the HSG clamp (60).
5. Remove the two Torx screws (10) that secure the top of the temperature module or
empty housing.
6. Remove the temperature module or empty housing.
Note
7. Complete the following steps that apply to your monitor:
You can remove and replace the temperature module without
disconnecting the main harness, battery harnesses, USB cables, or light-bar
harness from the main board.
If you are replacing only the temperature module without
replacing the temperature mounting cover, stop here. Install
the replacement module in the reverse order of disassembly
procedure.
If your monitor has a temperature module, remove the temperature mounting
cover:
Section A 113