252 Appendices
The following agents are compatible with the monitor. Follow the cleaning agent
manufacturer's guidelines:
70 percent isopropyl alcohol
10 percent chlorine bleach solution
Note
Note
Note
WARNING Liquids can damage electronics inside the monitor. Take care
to prevent water or other liquids from spilling on the monitor.
If liquids are spilled on the monitor:
1. Power down the monitor.
2. Disconnect the power plug.
3. Remove the battery pack from the monitor.
4. Dry off excess liquid from the monitor.
If liquids possibly entered the monitor, remove the monitor
Note
from use until it has been properly dried, inspected, and
tested by qualified service personnel.
5. Reinstall the battery pack.
6. Power on the monitor and verify that the monitor functions normally before
using it.
If liquids enter the printer housing:
1. Power down the monitor.
2. Disconnect the power plug.
3. Remove the battery pack from the monitor.
4. Remove and discard the paper roll.
5. Clean and dry the inside of the printer housing.
Note
The printer housing has a drain tube that directs liquids down
and out the bottom of the device. If liquids possibly entered
other openings in the monitor, remove the monitor from use
until it has been properly dried, inspected, and tested by
qualified service personnel.
6. Install a new roll of paper.
7. Power on the monitor and verify that the monitor functions normally before
using it.
CAUTION DO NOT use steam, heat, or gas sterilization on the monitor.
CAUTION DO NOT use harsh solvents such as acetone on the monitor.
Disinfect according to your facility's protocols and standards or local
regulations.
When cleaning the device, avoid using cloths or solutions that include
quaternary ammonium compounds (ammonium chlorides) or
glutaraldehyde-based disinfectants.
Disinfect according to your facility's protocols and standards or local
regulations.
®