114 Section A
Reassembly notes
Before installing the temperature module into the front housing, ensure that the
probe well seal is installed with the tab aligned with the slot or notch in the probe
well housing.
Route the power button flex cable up and over the temperature module and plug the
cable into the J6 ZIF locking connector on the main board.
When replacing the temperature module
1. Connect the USB cable to the temperature module.
2. Position the retaining clip on the USB cable against the shoulder of the mini-B
connector with the adhesive pad facing away from the temperature module and
toward the bottom.
3. Remove the backing on the retaining clip to expose the adhesive.
4. Rotate the clip on the USB cable to fasten the clip to the temperature module
housing.
a. Remove the screw that secures the temperature mounting cover to the
front housing.
b. Remove the temperature mounting cover (56) from the front housing.
If your monitor does not have a temperature module, remove the temperature
blanking panel:
a. Remove the two screws that secure the blanking panel to the front housing.
b. Remove the blanking panel (44).
®