168 Section B
1. Remove the two Torx screws that secure the printer housing to the top housing.
2. Remove the printer from the top housing:
a. Hold the printer housing while opening the printer door.
b. Separate the printer and printer door from the top housing.
Reassembly notes
1. Verify that the ground wire exits from the side opposite the printer board in the
notched area.
2. Place the printer door into the top housing assembly and hold it in place while
performing the next step.
3. Install the printer in the top housing by sliding the tabs into the slots.
4. Snap the printer door shut to hold the printer in place while securing it into the
housing with two screws.
5. Connect the printer ground wire to the bottom housing. (See the bottom housing
reassembly notes.)
When replacing the printer or adding a printer
If adding a printer, remove the printer door blank. Remove any remaining gasket
material and clean the surface with 70 percent isopropyl alcohol.
1. Separate the door from the printer assembly.
2. Follow the reassembly notes above to install the printer.
3. Connect the printer ground wire to the bottom housing. (See the bottom housing
reassembly notes.)
®