190 Functional verification and calibration
6. Select the device you want to test from the device list.
7. Click NIBP Sensor under the Device Information tab.
8. Click Leak Test in the NIBP Sensor pane on the right side of the window.
9. Follow the prompts until the test completes.
10. Click Close.
11. To view the results, open the active log file by selecting File > View Log File >
Active log file > Ok.
12. To record the results of your test, go to "Service record."
NIBP overpressure test (Standard unlicensed edition)
The NIBP overpressure test is performed automatically using the service tool. The
overpressure test verifies that the NIBP system will prevent the pressure from
exceeding 329 mmHg in adult mode and 164 mmHg in neonate mode. To pass this test,
the device must shut down the pump and open the valves when the pressure is
between 280 mmHg and 329 mmHg in adult mode, or 130 mmHg to 164 mmHg in
Neonate mode.
Perform an NIBP overpressure test
Note
1. Set up the test equipment.
2. Launch the service tool on the PC.
3. If the simplified interface is active, choose the Service option.
4. Log on with your user ID and password or with ADMIN as the user ID and leave the
password field blank.
5. Power on the device.
6. Select the device you want to test from the device list.
7. Click NIBP Sensor under the Device Information tab.
8. In the NIBP Sensor pane on the right side of the window, click Over Pressure Test.
9. Follow the prompts until the test completes.
10. Click Close.
11. To view the results, open the active log file by selecting File > View Log File >
Active log file > Ok.
12. To record the results of your test, go to "Service record."
NIBP accuracy check (Standard unlicensed edition)
The NIBP accuracy check is performed manually using the service tool to control the
valves. The accuracy check compares the reading from the primary transducer pressure
shown in the service tool window with the reading from an external calibrated digital
pressure meter. The results of this check are not recorded in the service tool log file. To
record the results for your records, copy the table in the service record section of the
service manual. If a calibration is required, run the NIBP calibration included with the
Gold licensed edition of the service tool.
If you are performing this test after performing the NIBP leak test, skip to
step 7.
®