108 Section A
•
When installing the housing top extension (75), verify that the top housing gasket is
in place. Use a 5/16-inch socket and Torque driver calibrated for 12.0 inch-pound
±1.0 inch-pound to secure the standoffs (74).
When replacing the top housing
For models without a printer, install the printer door blank as follows:
1. Remove the paper backing from the gasket on the door blank.
2. Install the door blank into the top housing, pressing to secure the door blank to the
top housing.
Remove the printer
Before you begin, remove the following as described in this manual:
•
Battery
•
Rear housing
•
CO2 module or blank faceplate
•
EarlySense module
•
SpO2 module or blank faceplate
•
NIBP module
•
Bottom housing
•
Top housing
1. Remove the two Torx screws that secure the printer housing to the top housing.
2. Remove the printer from the top housing:
a. Hold the printer housing while opening the printer door.
b. Separate the printer and printer door from the top housing.
®
Welch Allyn Connex
Vital Signs Monitor 6000 Series™