164 Section B
9. Disconnect and then remove the following:
a. The fan cable on the main harness from J45 on the main board.
b. The communications power cable on the main harness from J49 on the main
c. The battery power harness from J29 on the main board.
d. The smart-battery harness from J34 on the main board.
e. The printer harness from J17 on the main board.
f.
10. Remove the two screws (14) that secure the handle insert and remove the insert.
11. Remove the four Torx screws that secure the handle.
12. Remove the Pro Gaff tape securing the printer harness (if present) and light bar
harness to the LCD frame. Note the location of the Pro Gaff tape for placement
during reassembly.
board.
The speaker cable on the main harness from J12 on the main board.
Note
Use extra care when disconnecting J12 to avoid pulling
the connector off the board.
®