Service manual
Reassembly notes
Note
With the label facing down, slide the fan into the housing on top of the power-supply
cover until the tab on the cover holds it in place.
Clip the fan harness into the power-supply cover so that the connector is oriented
toward the fan, as shown in the following figure.
Remove the power supply
Before you begin, remove the following as described in this manual:
Battery
Rear housing
1. Remove the four Torx screws (10) securing the power-supply cover.
2. Remove the power-supply cover (18).
3. Disconnect the following from the power supply:
CAUTION Ensure that the fan is properly oriented. Air must flow toward
the power supply.
If the fan is dirty, use canned air to remove dust. Do not bend the fan
blades or otherwise damage the fan. If the fan is noisy and cleaning does
not reduce the noise, replace the fan.
The (green) ground cable
The AC power harness (17)
Note
See "Remove the AC power harness" for instructions.