Welch Allyn Connex Vital Signs Monitor 6000 Series Service Manual page 297

Hide thumbs Also See for Connex Vital Signs Monitor 6000 Series:
Table of Contents

Advertisement

Service manual
ECG basic functional test
St
Reading
ep
5
Waveform displayed:
5
HR displayed:
5
RR displayed:
6
Heart rate:
6
Respiration rate:
6
Seconds to display HR:
6
Seconds to display RR:
10
ECG waveform printed:
13
V-Tach alarm visible:
13
V-Tach alarm audible:
14
V-Tach alarm visible:
14
V-Tach alarm audible:
16
Asystole alarm visible:
16
Asystole alarm audible:
17
Asystole alarm visible:
17
Asystole alarm audible:
19
V-Fib alarm visible:
19
V-Fib alarm audible:
20
V-Fib alarm visible:
20
V-Fib alarm audible:
Acceptance criteria
Yes
No
Device displays ECG waveform
Yes
No
Device displays heart rate
Yes
No
Device displays respiration rate
♥/min 72 ≤ Reading ≤ 78 (♥/min)
BPM 13 ≤ Reading ≤ 17 (BPM)
seconds 16 seconds (10s+5*60/75♥/min +2s)
seconds 32 seconds (10s+5*60/15♥/min +2s)
Yes
No
Device prints ECG waveform
Yes
No
V-Tach alarm is displayed on device
Home tab
Yes
No
Alarm audio is audible
Yes
No
Visual alarm indication is cleared
Yes
No
No audio alarm
Yes
No
Asystole alarm is displayed on device
Home tab
Yes
No
Alarm audio is audible
Yes
No
Visual alarm indication is cleared
Yes
No
No audio alarm
Yes
No
V-Fib alarm is displayed on device Home
tab
Yes
No
Alarm audio is audible
Yes
No
Visual alarm indication is cleared
Yes
No
No audio alarm
Appendices 291
Pass
Fail