58 Troubleshooting
Message
Respiration alarm limits changed.
Weight scale messages
Message
Weight scale not functional. Call
for service.
Printer messages
Message
Alarm
Low battery, unable to print; plug
into outlet.
Welch Allyn Connex
Possible cause
You are not using Lead II to view
respiration
Attempting to acquire ECG/Impedance
respiration
Respiration physiological alarm limits
were modified because the respiration
source changed
Possible cause
The weight scale is not operating properly
Possible cause
The monitor's battery voltage is too low to
support printing
®
Vital Signs Monitor 6000 Series™
Suggested action
Check the patient cable
connection to the ECG
module.
Perform functional test. If
the test fails, contact
Customer Care and replace
the ECG module if
necessary.
Ensure that you are
viewing Lead II for
respiration.
Check electrodes and
leads, and replace as
needed.
Check the patient cable
connection to the ECG
module.
Perform functional test. If
the test fails, contact
Customer Care and replace
the ECG module if
necessary.
Reset alarm limits.
Suggested action
Check the scale.
Use the service tool to check
connectivity with adapter.
Replace cables.
Replace adapter.
Suggested action
Connect the monitor to AC power
to recharge the battery.