Welch Allyn Connex Vital Signs Monitor 6000 Series Service Manual page 61

Hide thumbs Also See for Connex Vital Signs Monitor 6000 Series:
Table of Contents

Advertisement

Service manual
Temperature messages (Braun ThermoScan PRO)
Message
Alarm
Unable to detect temperature.
Retry measurement.
Thermometer might be docked
improperly.
Braun temperature not functional.
Call for service.
Possible cause
Measurement was not taken
Measurement recalled from memory
Loose or broken USB cable
Dead battery
Braun display is blank or battery icon shows one
bar
Braun displays "POS" error
Braun displays "Err" error
Ambient temperature is outside the allowed
operating range (50–104°F or 10–40°C) or
changes too rapidly
Thermometer is not properly docked
Thermometer has dirty contacts
Communication timeout
BDD transfer took longer than expected
POST did not complete or failed within the
expected time period
POST failure
Braun display is all icons or blank
Unsupported parameter
Sensor data definition error (occurs when the
proxy is unable to set parameters to the sensor
using local DDS)
Troubleshooting 55
Suggested action
Retry measurement.
Check USB connection and cable.
Replace as necessary.
Recharge or replace battery.
Change the probe cover to reset.
Restrict patient movement and
ensure that the positioning of the
probe is correct and remains
stable while taking new
temperature.
Wait 20 seconds until
thermometer turns off
automatically, then turn on again.
Ensure thermometer and patient
are in an environment for 30
minutes where the temperature
is between 50ºF and 104ºF or
10ºC and 40ºC.
Re-dock thermometer.
Clean dock and thermometer
contacts with 70 percent
isopropyl alcohol. See Cleaning
Instructions for Braun
ThermoScan PRO 6000.
Remove and replace battery to
reset, then return thermometer to
dock. If error persists, return for
service.