Service manual
Note
11. Install the clear housing cover onto the module rear housing, and press the cover
into place until the three plastic clips lock.
12. Apply the two new blank labels to the locations shown in the figure below.
Remove the EarlySense module
Before you begin, remove the following as described in this manual:
Battery
Rear housing
1. Disconnect the USB cable from the EarlySense module.
The exhaust tube runs along the back of the module and
gradually bends to the exhaust fitting. The inlet tube and
wire run around the molded feature and boss to the large
connector.
Section A 91