4. Remove the power-supply cover (18).
5. Use a soft tool such as a spudger to separate the foam block (24) from the inside of
the rear housing.
Reassembly notes
Route the antenna cable in the channel located at the top of the rear housing.
When replacing the radio board and antenna
1. If the antenna is removed, clean the rear housing and mounting surface with
isopropyl alcohol.
2. Attach the antenna cable to terminal G on the radio board. Attach the other end of
the cable to the antenna on terminal PG.
3. Peel the backing off the foam block (24) and affix the foam block to the rear housing.
Section A 79