Welch Allyn Connex Vital Signs Monitor 6000 Series Service Manual page 293

Hide thumbs Also See for Connex Vital Signs Monitor 6000 Series:
Table of Contents

Advertisement

Service manual
Masimo SpO2, SpHb, heart rate, and RRa tests
Test
SpO2 heart rate
61 bpm
SpO2 saturation
81%
SpHb g/dL
14 g/dL
RRa
Nellcor SpO2 and heart rate test
Test
SpO2 heart rate
60 bpm
SpO2 saturation
75%
SpO2 heart rate
200 bpm
SpO2 saturation
90%
Specification
Pulse rate 61 ±1 bpm
60 to 62 bpm
Saturation 81% ±3%
78 to 84%
SpHb 14 g/dL ±1 g/dL
13 g/dL to 15 g/dL
A measurement between 5 and 16
breaths per minute
Specification
Pulse rate 60 ±1 bpm
59 to 61 bpm
Saturation 75% ±1 sat point
74 to 76%
Pulse rate 200 ±2 bpm
198 to 202 bpm
Saturation 90% ±1 sat point
89 to 91%
Appendices 287
Actual reading
bpm
%
g/dL
breaths per
minute
Actual reading
bpm
%
bpm
%
Pass
Fail
Pass
Fail