106 Section A
Plug the USB cables into the main board as follows:
Place the USB cables in the cable separator.
Note
Note
Note
Position each cable to extend approximately 1.5 inches past the separator.
Place the separator in the unit as shown below.
Plug the USB cables into the main board as follows:
If the USB cable was not removed from the temperature
module, leave the J1 position in the figure below vacant until
the cable separator is placed in the case and the other USB
cables have been connected to the main board. When the
cable separator is in the case, connect the USB cable from
the temperature module to J1 on the main board and snap
the cable into the J1 position in the cable separator.
The cable with right-angle connectors at both ends is in the
J14 position in the figure below.
If the option is not included, leave the separator position
empty.
®