180 Section B
Reassembly notes
Verify that the LCD flex cable feeds through the bezel's clearance feature. Insert the
LCD into the bezel in the front housing with the LCD flex cable on the lower right.
Position the display ground wire just below the LCD flex cable.
Place the LCD frame over the LCD. Verify that the LCD frame does not cover the
LCD flex cable.
When replacing the LCD frame
Install the LCD frame with shoulder screws and washers.
Install a 3-inch strip of Pro Gaff tape over the top part of the LCD harness cutout to
protect the LCD harness from abrasion.
Install the LCD harness. If it was removed from the frame, use isopropyl alcohol to
clean the area in the scribed boxes where you will mount the harness's ferrite
beads. Expose the sticky backing on the ferrite beads and mount them in the scribed
rectangles.
Note
Attach a piece of tape to the top of the LCD harness to secure it to the frame.
CAUTION The solder joint of the LCD connector (J48) is fragile. When
inserting or removing the flex cable, do not stress the solder joint.
If you don't place the ferrites correctly in marked areas, they
will obstruct the placement of other components like the
MCE board.
®