4. Cut the tie wrap that secures the main harness to the bottom bracket. This tie wrap
is located below the tie wrap on the main harness.
5. Pull the speaker wire out from behind the bottom bracket and disconnect the
speaker cable (26) from the main harness.
6. Disconnect the following on the battery connector board:
a. The battery power harness from J2.
b. The smart-battery harness from J3.
7. Remove the Pro Gaff tape securing the Ethernet cable and the LCD ground wire to
the metal chassis.
Note
Observe the locations of the Pro Gaff tape so that you can
replace it properly during reassembly.