Motor Replacement

Thermo Scientific
Use the following procedure to replace the motor.
Equipment required:
Hex drive, 5/32 and 5/64
Phillips drive, #2
1. Remove optical bench. Refer to
30.
2. Remove purge wheel housing. Refer to
Removal" on page 7-34.
3. Remove filter wheel. Refer to
35.
4. Using 5/64 hex drive, unfasten two #10-32 socket cap head screws.
Figure 7–41. Motor Removal from Motor Mount pt 1
"Optical Bench
Removal" on page
"Purge Wheel Housing
"Filter Wheel
Replacement" on page