12 Maintenance
12.3 Maintaining the reusable breathing
Requirements
Result

12.4 Replacing the hygiene filter

140
EN
MEDUVENT Standard
circuit
The reusable breathing circuit has been disassembled
"7.3.1 Disassembling the reusable breathing circuit", page
1. Check all parts of the reusable breathing circuit for external
damage, and to ensure that labeling is complete.
2. If necessary: Replace damaged or incorrectly labeled parts.
3. Replace PEEP control diaphragm and check valve diaphragm
(maintenance kit WM 17937).
4. Assemble the reusable breathing circuit
"7.3.2 Assembling the reusable breathing circuit", page
5. Punch the scheduled time for the next maintenance into the
service label (maintenance kit WM 17937).
6. Attach the service label to the device end of the ventilation
hose.
7. Perform a function check
The reusable breathing circuit has been maintained and is ready for
use.
Risk of injury due to contaminated or damaged hygiene filter!
A hygiene filter which is damaged or has been contaminated by a
prior infection transport might cause the patient and the user
serious or life-threatening injury.
 Check the hygiene filter and filter fleece for external signs of
damage and do not use if damaged.
 Replace damaged hygiene filter.
 Replace hygiene filter after every infection transport.
(see
(see "5 Function check", page
(see
74).
77).
45).