11 Maintenance
Result

11.3 Replacing the hygiene filter

124
EN
MEDUVENT Standard
4. Assemble the reusable breathing circuit
the reusable breathing circuit", page
5. Punch the scheduled time for the next maintenance into the
service label (maintenance set WM 15779).
6. Attach the service label to the device end of the ventilation
hose.
7. Perform a function check
The reusable breathing circuit has been maintained and is ready
for use.
Risk of injury due to contaminated or damaged hygiene filter!
A hygiene filter which is damaged or has been contaminated by
a prior infection transport might cause the patient and the user
serious or life-threatening injury.
⇒ Check the hygiene filter and filter fleece for external signs of
damage and do not use if damaged.
⇒ Replace damaged hygiene filter.
⇒ Replace hygiene filter after every infection transport.
Infection of the user or of the next patient resulting from
incorrect handling of a contaminated hygiene filter!
A contaminated hygiene filter might cause the patient or user to
suffer serious or life-threatening injury.
⇒ Always wear suitable personal protective equipment when
removing a contaminated hygiene filter.
⇒ Dispose of a contaminated hygiene filter when carrying out
hygienic reprocessing and do not reuse it.
(see "8.3.2 Assembling
90).
(see "5 Function check", page
44).