6. Repair information and repair instructions

6.1 General

Repairs to MEDUMAT Basic / Basic p should be carried out only at an ESD workstation!
Please follow the safety instructions for
MEDUMAT Basic / Basic p on page 6 of the
description and operating instructions.
All handling of the device pre-supposes a
precise knowledge of and compliance with
the description and operating instructions and
the service and repair instructions.
Please carry out only the repairs described in
these service and repair instructions. Other-
wise, perfect functioning of the
MEDUMAT Basic / Basic p cannot be guar-
anteed.

6.2 Replacing the sieve in the compressed gas connection

Tools required:
Slotted screwdriver,
Tweezers.
1. Unscrew the slotted screw at the compressed
gas connection
.
10
2. Using the tweezers, remove the sieve set
3. Carefully insert a new sieve set
compressed gas connection.
4. Screw the slotted screw back into the com-
pressed gas connection.
.
62
into the
62
Please ensure that your hands and workplace
are clean when carrying out repairs.
After each repair, please perform a functional
check (see "3. Final Check" on page 8).
When you replace components or individual
parts, please use original Weinmann parts only.
When ordering the housing base section
please specify the device model, year of con-
struction and device number.
Note:
The item numbers used in the following text
match the item numbers in the spare parts list
on page 47 and the overview on page 4.
62
Repair information and repair instructions
,
28
21