Connect Mobile Power Supply - Weinmann VENTIlogic LS Instructions For Use Manual

Ventilogic ls - single patient circuit with patient valve; ventilogic plus - single patient circuit with patient valve
Hide thumbs Also See for VENTIlogic LS:
Table of Contents

Advertisement

Available languages

Available languages

3
3

4.8 Connect mobile power supply

VENTIpower is available for power supply-independent operation. VENTIpower facilitates a
mobile power supply.
Warning!
• VENTIpower must be switched on at the On/Off switch before the start of
therapy so that in the event of a power failure or the failure of the internal
battery, VENTIpower can assume supply to the therapy device.
• Do not use VENTIpower as an uninterrupted power supply. VENTIpower does
not switch on automatically if the power supply fails. This can lead to the
patient being put in danger, especially in the case of life-support ventilation.
46
EN
Set up device
2. Plug the oxygen sensor (1) and air management
adapter into the T-adapter (3).
1
3. Plug the T-adapter (3) onto the device outlet port (5).
5
6
4. Connect the sensor (1) to the oxygen measuring jack
(4) with the aid of the cable.
5. Connect the patient circuit - with a bacteria filter if
required - as shown in the illustration.
6. Calibrate the oxygen sensor (see "7.3 Calibrate
oxygen sensor" on page 75).
4
If you want to use the therapy device with VENTIpower,
proceed as follows:
1. Connect VENTIpower to the therapy device.
2. Press the On/Off key to switch on VENTIpower.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ventilogic plus

Table of Contents