Solvent Information
Observe the following recommendations on the use of solvents.
• Follow recommendations for avoiding the growth of algae, see pump
• Small particles can permanently block capillaries and valves. Therefore,
• Avoid or minimize the use of solvents that may corrode parts in the flow
Recommended Wash Solvents
• water
• ethanol
• methanol
• water/acid (especially for basic compounds)
• water/base (especially for acidic compounds)
• water/acetonitrile
For different wash solvents as mentioned above, verify that the wash solvent is suitable
N O T E
for the silicone wash tubing.
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
manuals.
always filter solvents through 0.22 μm filters.
path. Consider specifications for the pH range given for different materials
like flow cells, valve materials etc. and recommendations in subsequent
sections.
Preparing the Fraction Collector
Solvent Information
4
63